A high-resolution diffractometer with low- and high-temperature cameras additionally equipped with a differential scanning calorimetry system for studying phase transformations within the temperature range of 93 K to 1773 K.

 

Technical charachteristics

  • • CuKα X-ray sealed tube (focus size – 0.4 x 12 mm).
  • • Rated operation X-ray sources: 40 kV/35 mA.
  • • Vertical Theta / Theta goniometer radius of 285 mm:
  •     • 2θ angle ranges from -3 to +162°;
  •     • scanning method – related θs / θd or θs, θd independent;
  •     • min scanning step value 0.0001°.
  • • High Energy-detector of the reflected X-rays DTEX / ULTRA:
  •     • 512 recording channels;
  •     • dynamic range of at least 105 pulses / s per channel.
  • • Optical System:
  •     • automatically varied gap on the output beam;
  •     • automatically varied gap on the diffracted beam;
  •     • automatically varied entrance slit;
  •     • automatic adjustment of the tube height, goniometer, optics and detector.
  • • Medium and low-temperature chamber Rigaku "R-300":
  •     • resistant, external to the sample, type of heater;
  •     • sample size of at least 17 x 25 x 2.0;
  •     • independent thermocouple and heater design;
  •     • temperature gradient across the sample does not exceed ± 2.5°C for temperatures not exceeding 100°C;
  •     • operating temperature in a vacuum ranges from -180°C to 300°C;
  •     • material of the sample holder – Cu.
  • • High-temperature chamber Rigaku "SHT-1500":
  •     • resistant, external to the sample, type of heater;
  •     • sample size of at least 11 x 25 x 0.5 mm;
  •     • independent thermocouple and heater design;
  •     • temperature gradient across the sample does not exceed ±0.5°C for up to 500°C temperatures and ± 5°C for above 1400°C temperatures;
  •     • max operating temperature in vacuum of at least 1450°C;
  •     • max operating temperature in the air at least 1500°C;
  •     • max operating temperature in an inert gas (He) is not less than 1300°C;
  •     • material of the sample holder – Pt.

Software

The PDXL software package provides an opportunity for a different analysis, for example, automatic qualitative phase analysis, quantitative phase analysis, refining unit cell parameters.

Information from the manufacturer

http://www.rigaku.com/products/xrd/ultima

Location

The Vasilyevsky Island training and research Complex of SPbSU.

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.