The Quantus fluorimeter is an instrument for measuring the concentration of nucleic acids. The device is used with a set of reagents Promega QuantiFluor® Dye Systems.

Before using the appliance for the first time, you need to calibrate the device, also you should calibrate before you start using each new set of reagents.

Materials:

0.5 PCR tube (thin-walled PCR tubes Axygen PCR-05-C);

20X TE ((pH 7.5) buffer;

dsDNA QuantiFluor®/ RNA QuantiFluor®.

Measurement of concentration of samples

  1. 1. In 0.5 PCR test tubes add 100 µl of QuantiFluor®, the necessary volume of samples (1-100 µl) and to increase the volume in the tubes to 200 µl of 1X TE.
  2. 2. Select the necessary Protocol, the volume of the analyzed sample and unit of concentration on the instrument display.
  3. 3. Put the sample into the instrument and close the lid. The concentration of the sample is shown in the display.

Detail: https://worldwide.promega.com/resources/protocols/technical-manuals/101/quantus-fluorometer-operating-manual-protocol/.

We use cookies on our website. Some of them are essential for the operation of the site, while others help us to improve this site and the user experience (tracking cookies). You can decide for yourself whether you want to allow cookies or not. Please note that if you reject them, you may not be able to use all the functionalities of the site.