Automated system ASAP 2020MP (Micromeritics) is used to analyze the surface area and pore structure study of dispersed materials. Allows full analysis of micro-and mesopores, their size distribution and surface area determination, warmth and adsorption rate. Measuring the pore diameter range from 0.35 to 500 nm. Samples can be analyzed with a specific surface area of 0.2 m2.
1. Measuring range of pore diameter: the lower limit of the range 0.35 nm , the upper limit of the range of 500 nm .
2. Studies of the physical and chemical adsorption including corrosive gases.
3. Two independent vacuum systems:
- - analysis of the sample,
- - to prepare a sample for analysis.
4. Pressure measuring range: lower limit of 0.001 mm Hg. Art., the upper limit of 950 mm Hg., resolution 0.001 mm. Hg. Art. measurement error not exceeding 0.15% of the measured value.
5. Integrated degassing system with two ports.
6. Temperatures degassing system: the lower limit of the range is not more than the ambient temperature, the upper limit of the range of 450oC .
7. Pressure range: lower limit range of 0 mm Hg. Art., the upper limit of the range of 950 mm Hg. Art.